72
of 216
Parameters in the "Channel settings" menu (M21)
Axis
X, Y
Active measurement channel, X or Y.
You cannot edit this parameter.
Socket
A, B, t (time)
Assign a physical connector to measurement channel X
or Y.
(Alternatively you can set a time axis here.)
Sensor
Assign the relevant sensor type:
Connector A: potentiometer or standard signal
Connector B: strain gage or standard signal
(Piezoelectric as an option)
Config.
Submenu for sensor-specific channel settings
6.3.1.1 Scaling analog sensors (strain gage, potentiometer, standard
signal sensors)
2-point scaling is used to associate the measured electrical signals with the measured physical quantities.
In this process, you assign a lower and upper calibration value (electrical quantity) to a lower and upper
scale value for the measured physical quantity. You can enter the "lower calibration" and "upper
calibration" values (electrical quantities) as numerical values or you can measure them using the
functions
[Teach in lower calibration]
and
[Teach in upper calibration]
.
Diagram 32: Scaling analog sensors