6.3.5.8 Graphical test operation – Reference curve
You can use
[Ref-Curve]
in the "Graphical test operation" menu (M59) to save the current measurement
curve as the reference curve and display this curve in the graph in measurement mode. The reference
curve is shown in violet and can be used by an operator to identify any differences between the actual
curve and ideal curve. A fewer number of measurement value pairs are saved for the reference curve.
1
In measurement mode, tap anywhere on the touchscreen. The icon appears in the
bottom-right corner.
2
Tap to open the "Configuration Main Menu".
3
Tap the "Program Setup" icon.
4
Tap the "Graph. TestOp." icon.
5
Tap
[Ref-Curve]
.
6
If the reference curve is
[Off]
, tap
[Off]
to
enable the function. Tap
[Save]
to save the
displayed measurement curve as the
reference curve.
Note:
This will overwrite any reference
curve that you may have saved
previously.