8.11.14 Entering the periodicity of a maintenance operation to be
carried out on the sensor connected to the analogue input
Refer to section 8.9 to access the "Calibration" menu.
Calibration intervals
Maintenance
ENTERING
Reset Mainte-
nance Timer
Yes/No
Calibration
Mx:Inputs
AI1/AI2
Mode:general
Mode:chlorine
On each due date, the 8619 generates a "maintenance" event, signalled on the display by the icon , and a
"warning" event, signalled on the display by the icon
.
→
Enter the number of days between two maintenance reminders.
When the maintenance operation has been performed, you can choose ("Yes") or not ("No") to restart the
countdown of days until the next maintenance operation in the menu "Reset maintenance Timer".
→
Select "Reset Maintenance Timer" -> "Yes" or -> "No"
→
Enter "0000 days" to disable the automatic reminder for the maintenance operation.
8.11.15 Restoring the factory calibration of the analogue inputs
Refer to section 8.9 to access the "Calibration" menu.
Yes/No
Factory calibr.
Calibration
Mx:Inputs
AI1/AI2
Mode:general
Mode:chlorine
8.11.16 Resetting the totalisers
This function is available on the devices with analysis modules if the software option, "FLOW", is activated. See
section 8.10.4
Refer to section 8.9 to access the "Calibration" menu.
Calibration
DI1/2 totaliser
Reset totalis. A
M0:Inputs
Reset totalis. B
Yes/No
Yes/No
Mx:Inputs
141
Adjustmentandcommissioning
Type 8619
English
Summary of Contents for 8619 multiCELL WM AC
Page 6: ...6 English ...
Page 12: ...12 Type 8619 English ...
Page 24: ...24 Type 8619 English ...
Page 40: ...40 Type 8619 English ...
Page 70: ...70 Installation and wiring Type 8619 English ...
Page 180: ...180 Type 8619 English ...
Page 192: ...192 Type 8619 English ...
Page 193: ......
Page 194: ...www burkert com ...