60
Adjustmentandstart-up
9.14.2 checking the outputs functions
See chpt. 9.9 to access the Test menu.
• Make sure that the “Hold” mode is deactivated. Refer to chpt. 9.12.1.
• The
T
icon is displayed in place of the
icon as soon as the check for the correct working of an
output has started. During the check the related output does not react according to the measured
physical value.
Test
Outputs
AC2:
AC1:
INPUT
INPUT
TR1:
OFF/ON
TR2:
OFF/ON
AC1
: check that current output 1 is working correctly by entering a current value and then selecting “OK”.
AC2
: check that current output 2 is working correctly by entering a current value and then selecting “OK”.
TR1
: check that transistor output 1 is working correctly by selecting the status of the transistor (“ON” or “OFF”)
then “OK”.
TR2
: check that transistor output 2 is working correctly by selecting the status of the transistor (“ON” or “OFF”)
then “OK”.
9.14.3 checking the outputs behaviour
See chpt. 9.9 to access the Test menu.
• Make sure that the “Hold” mode is deactivated. Refer to chpt. 9.12.1.
• The
T
icon is displayed in place of the
icon as soon as the check for the correct working of an
output has started. During the check the related output does not react according to the measured
physical value.
The feature allows for simulating the measurement of the process value to check if the outputs are correctly
configured.
Test
PVar:
Value:
Sensor
INPUT
CondS
TDSppm
TempC
TempF
CondR
PVAR
: choose the process value to be tested.
VALUE
: enter a process value selected from the “PVAR” function above to check output behaviour.
English
Type 8222 ELEMENT