Adjustmentandstart-up
9.12.5 Entering an offset for the temperature measurement
See chpt. 9.9 to access the Calibration menu.
The temperature transmitted by the Pt1000 probe may be corrected. This correction value is the temperature
offset.
Calib
Sensor
Temperature
INPUT
Enter the temperature offset
(input limits: ±5 °C)
9.13 Knowing the Diagnostic menu
9.13.1 Modifying the Diagnostic menu access code
See chpt. 9.9 to access the Diagnostic menu.
Code
0***
Confirm code
0***
Diagnostic
System
Enter the new
code
Confirm the new
code
If the default code (0000) is entered, the code will not be requested to access the menu.
9.13.2 Monitoring the fluid conductivity
See chpt. 9.9 to access the Diagnostic menu.
The function allows for monitoring the fluid conductivity and configure the behaviour of the device if the param-
etered ranges are exceeded.
A malfunction in your process or the conductivity sensor may be indicated either by too low or too high a fluid
conductivity or by an incorrect conductivity measurement.
Diagnostic
Conductivity
Sensor
Conductivity
Activate:
Yes/No
READ
Warn lo:
INPUT
Err hi:
INPUT
Err lo:
INPUT
Warn hi:
INPUT
To be warned when the fluid conductivity is too low or too high:
→
activate monitoring of the fluid conductivity in the function “activate”, then
→
set a conductivity range outside of which the device generates a “warning” event and displays the
and
icons.
→
set a conductivity range outside of which the device generates an “error” event and displays the
and
ERR
icons.