178
Maintenance and troubleshooting
Type 8098
19.8.23
Message "Totalizer 1 started" or "Totalizer 2 started"
Message
Totalizer 1 started
or
Totalizer 2 started
Symbol displayed
-
Possible cause
The related totalizer has been started by the user.
What to do?
-
19.9
Messages due to calibration or simulation
19.9.1
Message "Calibration result out of range"
Message
Calibration result out of range
Symbol displayed
(symbol changed)
Possible cause
The calibration has failed. The calibration has failed because of 1 of the
possible causes:
•
An event such as
Change of liquid
or
Sound cond. out of range
has
been generated during the calibration.
• The calculated offset of the density factor is lower than 0.5 or higher than
2.
• The calculated offset of the acoustic transmission factor is lower than 0.5
or higher than 2.
What to do?
→
Make sure the liquid is the same during the calibration procedure.
→
Make sure the conditions are met to measure the flow rate correctly.
→
Do a new calibration.
19.9.2
Message "Zero calibration cancelled, the flow rate is higher
than 5% of full scale"
Message
Zero calibration cancelled, the flow rate is higher than 5% of full scale
Symbol displayed
Possible cause
The calibration has failed because the
flow rate is less than 5% of the full
scale.
What to do?
→
Make sure the flow is stopped in the pipe.
→
Do a new calibration.