Outputs – Maintenance
Type 8098 FLOWave L
20
OUTPUTS – MAINTENANCE
The settings can be made with the
Installer
user level.
20.1
Calibrating an analogue output
The analogue outputs are calibrated at the factory.
To adjust the analogue output to your equipment, do the following:
→
Connect a multimeter to the analogue output you want to adjust.
→
Go to the
CONFIGURATION
view.
→
Outputs
→
Confirm to access the
Parameter
view.
→
Go to the
MAINTENANCE
view.
→
Output 1: analog
or
Output 3: analog
→
Calibration
The
Current settings
are displayed
The device gen-
erates a 4 mA current on the chosen analogue output.
→
Enter the current value measured by the multimeter
The device generates a
20 mA current on the chosen analogue output.
→
Enter the current value measured by the multimeter
The
New settings
are
displayed.
→
Save.
The analogue output is adjusted.
20.2
Checking the correct operation of an analogue
output
To check the correct operation of an analogue output, do the following:
→
Connect a multimeter to the analogue output you have adjusted.
→
Go to the
CONFIGURATION
view.
→
Outputs
→
Confirm to access the
Parameter
view.
→
Go to the
MAINTENANCE
view.
→
Output 1: analog
or
Output 3: analog
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...