282
SAW sensor – Maintenance
Type 8098 FLOWave L
17.26
Resetting all the calibration data to its default values
(additional measurement values)
The calibration data that can be reset is:
•
the offset value of the DF,
• the slope value of the DF,
•
the offset value of the liquid density,
• the slope value of the liquid density,
•
the offset value of the liquid concentration,
• the slope value of the liquid concentration,
•
the offset value of the acoustic transmision factor,
• the slope value of the acoustic transmision factor.
To reset all the calibration data to its default values, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Go to the
MAINTENANCE
view.
→
Calibration
→
Add. meas. values
→
Reset to default
→
Confirm.
All the calibration data is reset to their default values.
→
Go back to the parent menu.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...