270
SAW sensor – Maintenance
Type 8098 FLOWave L
17.9
Resetting the flow rate calibration data to its default
values
To reset all the flow rate calibration data to its default values, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Go to the
MAINTENANCE
view.
→
Calibration
→
Stand. meas. values
→
Flow rate
→
Reset to default
Confirm.
All the flow rate calibration data is reset to their default values.
→
Go back to the parent menu.
17.10
Setting the offset value of the liquid temperature
Instead of setting the offset value of the liquid temperature, you can calibrate it. See chapter 17.11 Cali
brating the offset value of the liquid temperature
To enter an offset value for the liquid temperature, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Go to the
MAINTENANCE
view.
→
Calibration
→
Stand. meas. values
→
Temperature
→
Offset
Enter value
→
Set the value of the offset.
→
Save.
The offset value of the liquid temperature is set.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...