SAW sensor – Maintenance
Type 8098 FLOWave L
→
Stand. meas. values
→
Flow rate
→
Offset
→
Zero flow offset by teach-in
The
Current settings
are displayed.
→
Stop the flow and wait until it is completely still.
Start the calibration of the offset value.
After 30 s, the
New settings
are displayed.
→
Save.
The offset value of the flow zero point is calibrated.
If the calibration fails, a message is displayed. Refer to chapter 24.10 Messages due to calibration or
→
Acknowledge the message to go back to the parent menu.
17.6
Setting the offset value of the flow zero point
Instead of setting the offset value of the flow zero point, you can calibrate it. See chapter 17.5 Calibrating
the offset value of the flow zero point
.
To enter the offset value of the flow zero point, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Go to the
MAINTENANCE
view.
→
Calibration
→
Stand. meas. values
→
Flow rate
→
Offset
→
Enter value
→
Set the value of the offset. Take into account the direction of the flow as set in chapter 17.4
Setting the direction of the flow
→
Save.
The offset value of the flow zero point is set.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...