241
SAW sensor – Parameter
Type 8098 FLOWave L
15.15
Getting as accurate measurements of the volume
flow rate, the mass flow rate or the liquid velocity as
possible
To get as accurate measurements of the volume flow rate, the mass flow rate or the liquid velocity as pos
-
sible, you can activate the compensation of the kinematic viscosity (in mm
2
/s).
The following kinematic viscosity compensations are available:
• for water or a liquid whose viscosity
u
(in mm
2
/s) varies with the temperature
T
(in °C) like the viscosity of
water and in the same range as water. Default setting. The related equation is:
𝑣𝑣 =
1
0, 0,020217𝑇𝑇 + 9,9. 10
−5
𝑇𝑇
2
→
To activate the viscosity compensation for water, see chapter 15.15.1
• for a liquid with a constant viscosity. To be chosen if the liquid temperature is constant and thus the vis-
cosity of the liquid is constant. The related equation is:
𝑣𝑣
=
𝑎𝑎
→
To activate the viscosity compensation for a liquid whose viscosity is constant, see chapter 15.15.2
.
• for a liquid with a linear compensation curve. To be chosen if the viscosity of the liquid varies in a linear
way depending on the liquid temperature. The related equation is:
𝑣𝑣 = 𝑎𝑎 + 𝑏𝑏𝑏𝑏
→
To activate the viscosity compensation for a liquid with a linear viscosity compensation curve, see
chapter 15.15.3
.
• for a liquid with a quadratic compensation curve. To be chosen if the viscosity of the liquid varies in a
quadratic way depending on the liquid temperature. The related equation is:
𝑣𝑣 = 𝑎𝑎 + 𝑏𝑏𝑏𝑏 + 𝑐𝑐𝑏𝑏
2
→
To activate the viscosity compensation for a liquid with a quadratic viscosity compensation curve, see
chapter 15.15.4.
• for a liquid with an inverse quadratic compensation curve. To be chosen if the viscosity of the liquid varies
in an inverse quadratic way depending on the liquid temperature, but the viscosity range is different from
the one of water. The related equation is:
𝑣𝑣 =
1
𝑎𝑎 + 𝑏𝑏𝑏𝑏 + 𝑐𝑐𝑏𝑏
2
→
To activate the viscosity compensation for a liquid with an inverse quadratic viscosity compensation curve,
see chapter 15.15.5
.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...