SAW sensor – Parameter
Type 8098 FLOWave L
→
Stand. meas. values
→
Density
→
Limits
→
Reset to default
→
Confirm.
The limit values and the hysteresis value are reset.
→
Go back to the parent menu.
15.9.9
Setting the measurement mode of liquid density
Density mode can be selected through the following values:
•
Measured (value measured via SAW signals, and liquid temperature)
•
Water (value calculated based on liquid temperature measured by FLOWave)
• Constant (value set to a constant value)
•
Linear (
r
= a+bT; coefficients a and b to be set by installer, T in °C)
•
Quadratic (
r
= a+bT+cT
2
; coefficients a, b and c to be set by installer, T in °C)
When option
Density
is ordered, default density mode will be
Measured
.
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Stand. meas. values
→
Density
→
Mode
The
Current settings
are displayed.
→
Choose the density mode between
Measured
,
Water
,
Constant
,
Linear
,
Quadratic
→
If coefficients have to be set (in
Constant
,
Linear
,
Quadratic
mode), then set the values
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...