132
General settings – Diagnostics
Type 8098 FLOWave L
13.3
Reading out data related to büS
13.3.1
Reading out the number of current receive errors
To read out the number of current receive errors, do the following:
→
Go to the
CONFIGURATION
view.
→
General settings
→
Confirm to access the
Parameter
view.
→
Go to the
DIAGNOSTICS
view.
→
büS status
→
Receive errors
→
Go back to the parent menu.
13.3.2
Reading out the maximum number of receive errors since the
last power-up of the device
To read out the maximum number of receive errors, do the following:
→
Go to the
CONFIGURATION
view.
→
General settings
→
Confirm to access the
Parameter
view.
→
Go to the
DIAGNOSTICS
view.
→
büS status
→
Receive errors max.
→
Go back to the parent menu.
13.3.3
Reading out the number of current transmit errors
To read out the number of current transmit errors, do the following:
→
Go to the
CONFIGURATION
view.
→
General settings
→
Confirm to access the
Parameter
view.
→
Go to the
DIAGNOSTICS
view.
→
büS status
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...