TC-Standard (+) X2
7.6 Drying of the moisture detector (option)
The moisture detector must be dried if moisture enters.
– Close the gas supply.
– Switch off and unplug the device.
– Loosen the swivel nut for the moisture detector connection line and disconnect the line.
– Unscrew the moisture detector counter-clockwise and remove.
– Dry moisture detector.
– Reinsert the moisture detector and carefully tighten the screw connection.
– Connect the connection line and tighten the swivel nut.
– Restore the power and gas supply.
7.7 Calibration of the moisture detector (option)
– When replacing the moisture detectors, they must be recalibrated.
– Be sure dry gas flows through the cooler.
– Select cooler menu and confirm.
or
– Select menu item moisture detector.
– The display shows (Reset).
– Confirm the display to calibrate the moisture detectors.
For a detailed overview of menu navigation, refer to chapter "Operation and Control".
7.8 Spare parts and accessories
Please also specify the model and serial number when ordering parts.
Upgrade and expansion parts can be found in our catalog.
Available spare parts:
Item no.
Description
91 00 10 00 07
Display module MCD400
91 44 05 00 80
Connecting cable controller board display module
91 00 11 01 80
Microcontroller board LPP MCP2
40 11 00 0
Flow adapter type G, PVDF G1/4
40 11 00 0I
Flow adapter type NPT, PVDF NPT 1/4"
41 11 10 0
Moisture detector FF-3-N, without cable
91 44 05 00 81
Moisture detector connection cable, 300 mm
91 44 05 00 82
Moisture detector connection cable, 450 mm
41 50 29 99
Filter AGF-PV-30-F2, G1/4
41 50 29 99 I
Filter AGF-PV-30-F2, NPT 1/4"
44 65 90 00 5
Fan, 12 V DC
91 00 01 01 98
Power board
91 00 01 11 98
Power board, 24 V DC
91 00 01 11 87
Controller board
see data sheet 450022
Peristaltic Pumps CPsingle, CPdouble X2
31
Bühler Technologies GmbH
BE440023 ◦ 10/2018