• An overcurrent protection device must be fitted as a disconnecting device in all supply lines but must not interrupt the pro-
tective conductor. If fuses are used, all must have the same rated value and the same tripping characteristics and must be
installed side by side. They are preferably to be placed before the power switch. Radio interference control equipment
between the AC input and overcurrent protection device is allowed. Fuses and single-pole circuit breakers are not built into
the neutral conductor of multi-phase devices.
If the above-cited isolating devices are already in the system, these no longer need to be provided by the operator for the system.
Please see the chapter ‘Technical Data’ for the applicable values of the isolating devices.
5 Operation and controls
NOTICE
The device must not be operated beyond its specifications.
5.1 Basic function of the probe controller
5.1.1 Controller function
After switching on the combination the probe is heated up. The display with the current temperature will light up on the regu-
lator. As long as the set operating range is has not yet been reached, the display will flash and the status contact is in Alarm posi-
tion. Once the working range has been reached, the status contact switches and the display is steady.
The target temperature, the working range of the probe and the temperature unit (°C/°F) are set using the three control buttons
on the controller. This is described in chapter “Operation and use”.
The factory settings are:
Unit °C
Sample probe temperature 280 °C, operating range ±10 °C
Heated adapter 160 °C, operating range ±10 °C
5.2 Use of menu functions
Overview of the operational principal:
Use this short description if you have experience with the device.
Operation is carried out by only the keys with the following functions:
Key
Function
• Switch from measurement display to main menu
• Selection of the display menu item
• Accepting the changed value or selection
• Switch to the upper menu item
• Increase of the value of switching the selection
• Temporary display of the alternative measurement display (if option is installed)
• Switch to lower menu item
• Decrease of the value of switching the selection
• Temporary display of the alternative measurement display (if option is installed)
27