BÜHLER BU3011 peristaltic pump

Safety information

Refer to the installation manual for general safety information, hazard descriptions and precautionary labels
descriptions.

Introduction

This document applies to sampler BÜHLER BU3011.XX.PXX51 with peristaltic pump. Refer to 

Figure 1

.

Figure 1  Sampler versions

1   

Sampler with dosing vessel

2   

Sampler with peristaltic pump

Calibrate the sampler

Make sure to calibrate the sampler before the sampler is put into operation. Calibration must be the first on-site
step each time a new sampling point is used.

Items to collect:

Measuring glass with 2 L minimum capacity

Measuring glass with 200 mL minimum capacity

1.

On the sampler menu, select 

SETUP

 > 

SYSTEM SETTINGS

 > 

CALIBRATION VOL

.

2.

Push 

NEXT

 to start the calibration.

3.

Select 

START UPPER VOLUME

. Push 

NEXT

.

The sampler automatically starts three calibration cycles.

4.

Use the 2 L measuring glass to collect the sample of the three calibration cycles.

5.

Enter the total volume of the three cycles. Push 

NEXT

.

The sampler shows the message "UPPER VOLUME CALIBRATED"

6.

Select 

START LOWER VOLUME

. Push 

NEXT

.

The sampler automatically starts three calibration cycles.