BÜHLER BU3011 peristaltic pump
Safety information
Refer to the installation manual for general safety information, hazard descriptions and precautionary labels
descriptions.
Introduction
This document applies to sampler BÜHLER BU3011.XX.PXX51 with peristaltic pump. Refer to
.
Figure 1 Sampler versions
1
Sampler with dosing vessel
2
Sampler with peristaltic pump
Calibrate the sampler
Make sure to calibrate the sampler before the sampler is put into operation. Calibration must be the first on-site
step each time a new sampling point is used.
Items to collect:
•
Measuring glass with 2 L minimum capacity
•
Measuring glass with 200 mL minimum capacity
1.
On the sampler menu, select
SETUP
>
SYSTEM SETTINGS
>
CALIBRATION VOL
.
2.
Push
NEXT
to start the calibration.
3.
Select
START UPPER VOLUME
. Push
NEXT
.
The sampler automatically starts three calibration cycles.
4.
Use the 2 L measuring glass to collect the sample of the three calibration cycles.
5.
Enter the total volume of the three cycles. Push
NEXT
.
The sampler shows the message "UPPER VOLUME CALIBRATED"
6.
Select
START LOWER VOLUME
. Push
NEXT
.
The sampler automatically starts three calibration cycles.