TC-Standard (+)
7.6 Drying of the moisture detector (option)
The moisture detector must be dried if moisture enters.
– Close the gas supply.
– Switch off and unplug the device.
– Loosen the swivel nut for the moisture detector connection line and disconnect the line.
– Unscrew the moisture detector counter-clockwise and remove.
– Dry moisture detector.
– Reinsert the moisture detector and carefully tighten the screw connection.
– Connect the connection line and tighten the swivel nut.
– Restore the power and gas supply.
7.7 Calibration of the moisture detector (option)
– When replacing the moisture detectors, they must be recalibrated.
– Be sure dry gas flows through the cooler.
– Select cooler menu and confirm.
or
– Select menu item moisture detector.
– The display shows (Reset).
– Confirm the display to calibrate the moisture detectors.
For a detailed overview of menu navigation, refer to chapter "Operation and Control".
7.8 Replacing the base unit
– Close gas supply.
– Switch off device and disconnect all plugs (e.g. status output connector, supply input, etc.).
– Remove the base unit.
– Remove options/add-ons.
– Install options on the new unit.
– Restore the power and gas supply.