TC-Standard (+)
lower alarm limit (alarm low)
Display → Cooler →
A Lo
Here you can set the lower threshold for the visual signal and the alarm relay. The alarm limit is set
based on the cooler temperature setting.
Parameter range:
-1 °C to -3 °C (-1.8 °F to -5.4 °F)
Factory setting:
-3 °C (-5.4 °F)
Note:
This menu item is hidden if the keylock is enabled.
DeltaT (dt)
Display → Cooler →
dt
Here you can set the nominal difference with respect to ambient temperature.
Parameter range:
-30 K…0 K
Factory setting:
-15 K
Note:
This menu will be hidden if the menu is locked.
DeltaT mode (Modu)
Display → Cooler →
Modu
Here you can activate or deactivate DeltaT-mode.
Parameter range:
AbSo
,
dt
Factory setting:
AbSo
(normal operation mode)
Note:
This menu will be hidden if the menu is locked.
Exit submenu 1
Display → Submenu →
E
Selecting this will return you to the main menu.
5.4.4 Submenu 1 (global settings)
Temperature unit
Display →
toP
→
unit
Used to select the temperature display unit.
Parameter range:
°C
,
°F
Factory setting:
°C
Analog output
Display →
toP
→
An1
This submenu is used to specify the settings for analog output 1, see chapter Submenu 2 (Analog
Output 1)
Note:
This menu will be hidden if the menu is locked.
21