5 Putting into operation
34
K-415 Operation Manual, Version B
5 .8
Electrical connections
Notice
Risk of instrument damage by wrong mains supply.
• External mains supply must meet the voltage given on the type plate.
• Check for proper grounding.
• Exchange defective cabling instantly.
• Make sure the used digester and the K-415 are suitable for the same power system (voltage
and frequency) when both instruments share the same power source.
Connect the Scrubber K-415 to the mains with the power cord contained in the scope of delivery.
To start the Scrubber automatically at the same time the digestion process starts, connect it to the
SpeedDigester K-439, to a Digest System K-437 or to a Digest Automat K-438/K-432 with the
optional connection cable.
NOTE
To cut the power in case of an emergency by unplugging, the instrument or any other item must not
block the mains plug! In this case, the plug must be able to be pulled out instantly.
5 .9
Preparing washing solutions
!
WARNING
Serious chemical burns by corrosives.
• Observe all data sheets of the used chemicals.
• Handle corrosives in well ventilated environments only.
• Always wear protective goggles.
• Always wear protective gloves.
• Always wear protective clothes.
• Do not use damaged glassware.
Summary of Contents for Scrubber K-415
Page 1: ...Scrubber K 415 Operation Manual 11593505 en ...
Page 5: ...1 About this manual 4 K 415 Operation Manual Version B ...
Page 7: ...2 Safety 6 K 415 Operation Manual Version B ...
Page 13: ...3 Technical data 12 K 415 Operation Manual Version B ...
Page 19: ...4 Description of function 18 K 415 Operation Manual Version B ...
Page 23: ...5 Putting into operation 22 K 415 Operation Manual Version B ...
Page 39: ...6 Operation 38 K 415 Operation Manual Version B ...
Page 49: ...9 Shutdown storage transport and disposal 48 K 415 Operation Manual Version B ...
Page 51: ...10 Spare parts 50 K 415 Operation Manual Version B ...
Page 55: ...11 Declarations and requirements 54 K 415 Operation Manual Version B ...