5 | Installation
BÜCHI Labortechnik AG
22/60
Operation Manual Rotavapor® R-100
5.3
Fitting the vapor duct and condenser
1
4
3
2
2
u
Insert vapor duct (4) into the rotary drive unit.
The vapor duct snaps into place with an audible click.
u
NOTICE! Risk of property damage due to vapor duct falling out if incorrectly
fitted.
Pull on the vapor duct to check that it is properly engaged and cannot be
pulled out without resistance.
u
NOTICE! Risk of seal damage if fitted incorrectly.
Insert seal (2) into the flange
of the condenser (1) as shown.
u
Fit condenser (1) into the rotary drive unit, keeping it straight as you do so. When
doing so, make sure that the inner lip of the seal (2) does not fold over as the seal
would then be destroyed.
u
Tighten the cap nut (3) to fix the condenser in position. When doing so, make sure
that the spring clip in the cap nut completely encircles the neck of the condenser.