6 | Operation
BÜCHI Labortechnik AG
36/70
Operation Manual Pure Chromatography Instrument C-805
Add steps to the Gradient graphic
u
Tap on the line at the time for which you want to edit the gradient and drag it to the
desired %B, then release.
Deleting steps from the Gradient graphic
u
To remove a step, drag the point to the baseline or to any gray area around the
graph until a red X is visible, then release.
6.2.4
Editing the wavelength for the UV detector
Status Checkbox
Explanation
Explanation
On
The instrument collects fractions.
Monitoring
The instrument records the data from the
UV detector but does not collect the frac-
tions.
Off
The instrument does not record data
from the UV detector and does not col-
lect fractions.
Navigation path
➔
Wavelength selection panel
Precondition:
R
The UV detector is selected.
u
Navigate to the
Wavelength
panel according the navigation path.
u
Activate the checkbock next to
[UV]
.
u
Tap the input box next to
[UV]
.
ð
The display shows a dialog box with a numeric input box.
u
Enter a value for the wavelength.
u
Tap the button
[OK]
.
ð
The dialog box closes.
ð
The wavelength is saved.