3 | Product description
NIR-Online GmbH
14/36
Operation Manual Process Analyser X-One
Rating
Meaning according to Directive 2014/34/EU
Ex
Explosion-safe
tb
Protection class protected by enclosure
[op is Da]
Intrinsically safe optical radiation approved for dust zone 20
IIIC
Dust group conductive dusts
T80 °C
Temperature classification max. surface temperature = 80° C
Db
Device safety level zone 21, adequate safety in event of fore-
seeable faults
3.5
Scope of delivery
NOTE
The scope of delivery depends of the configuration of the purchase order.
Accessories are delivered as per the purchase order, order confirmation, and deliv-
ery note.
3.6
Technical data
3.6.1
Sensor
Specifications
X-One
Dimensions (W x D x H) 220 x 220 x 135 mm
Weight
7.5 kg
Max. operating pressure 30 bar at flange
Product temperature
(temperature at flange
with water cooling)
-10 °C to +130 °C
Product temperature
(temperature at flange
without water cooling)
-10 °C to + 70 °C
Vibrations
0.2 G at 0.1 - 150 Hz
Wavelength spectrum
NIR range
900 - 1700 nm; 11100 - 5900 cm
-1
Detector
Diode array
Average measurement
time
V3: 50 spectra/s
V3S: 200 spectra/s
IP code
IP66 (IEC 60529)
Type of lamp
Tungsten-halogen dual lamp
Lifetime lamp
18000 h (2 x 9000 h)
Minimum clearance on
all sides
100 mm
Connection Voltage
85 to 264 VAC
Frequency
50/60 Hz