8 Troubleshooting
88
B-90
Operation Manual, Version B
Table 8-1: Error messages and their remedy (cont.)
011
Inlet temperature exceeded
Heating control or heating relais
defective
Switch off the device and try
again. If the error still occurs
replace the heater module or
contact the BUCHI customer
service.
014
Voltage HV-Module
HV-module defect or internal
wiring interruption or quality fall-
off voltage because of a short
circuit
Switch off the device and try
again. If the error still occurs
contact the BUCHI customer
service.
015
Current in HV-Module too high
HV-module defect or internal
wiring interruption
Switch off the device and try
again. If the error still occurs
contact the BUCHI customer
service.
016
Current in HV-Module too low
HV-connector not plugged in or
defective wiring error
Switch off the device and try
again. Check the HV-module,
connection. If the error still
occurs , contact the BUCHI
customer service.
020
Relative pressure sensor 1
not connected
Sensor not connected or defec-
tive internal wiring
Contact BUCHI customer service
021
Relative pressure sensor 2
not connected
Sensor not connected or defec-
tive internal wiring
Contact BUCHI customer service
023
Difference pressure sensor
not connected
Sensor not connected or defec-
tive internal wiring
Contact BUCHI customer service
BUCHI
024
Inside pressure low
System not leakproof
Test for leaks
025
Filter pressure drop low
Filter not installed or defective
internal wiring
Check if outlet filter is blocked.
Replace the filter.
026
Flow rate low
No flow rate, system not leak-
proof
Generate gas flow, test for leaks
027
Inside pressure high
Gas pressure connection too
high
Regulate gas pressure
028
Flow rate high
Sensor is at stop position
Reduce gas flow rate
029
Relative pressure sensor 1
defective
Sensor or internal wiring defec-
tive
Contact BUCHI customer service
030
Relative pressure sensor 2
defective
Sensor or internal wiring defec-
tive
Contact BUCHI customer service
032
Difference pressure sensor
defective
Sensor or internal wiring defec-
tive
Close gas flow rate and restart
the system, contact BUCHI
customer service
035
No glass isolation cylinder
Position switch
Install glass insolation cylinder
040
Valve 1 CO
2
not connected
Internal wiring defective or
missing
Contact BUCHI customer service
041
Valve 2 O
2
not connected
Internal wiring defective or
missing
Contact BUCHI customer service
042
O
2
sensor not connected
Defective sensor, sensor cable or
internal wiring
Contact BUCHI customer service