Installation | 5
Operation Manual Lyovapor™ L-300/L-300 Pro
27/118
Fig. 12: Castor brake
NOTE
Make sure that the power supply can be disconnected at any time in an emergency.
NOTE
Depending on the ambient conditions, condensation may form on any cold surface
on the unit.
5.2
Putting the instrument in operation
NOTICE
Instrument damaged if restarted too early
Wait ten minutes before restarting the instrument. The oil in the refrigerant
compressor requires ten minutes to return to the collection tank.
5.2.1
Preparing the instrument
u
Clean the instrument with a damp cloth before commissioning.
u
Check all sealing surfaces for scratches, dust, and cleanliness.
5.2.2
Establishing electrical connections
NOTE
Observe the regulatory provisions when connecting the Lyovapor™ to the power
supply.
u
Use additional electrical safety features (e.g., residual-current circuit breakers) to
comply with local laws and regulations.
The power supply must fulfil the following conditions:
1. Provide the mains voltage and frequency specified on the type plate of the
instrument.