10 | Help with faults
Büchi Labortechnik AG
92/100
Operation Manual Lyovapor™ L-200/L-200 Pro
Error message
Possible cause
Solution
5278
Pressure increases above
level of 1.5 mBar during leak
test
u
Resolve any leaks.
u
Clean and dry condenser
chamber.
u
Clean O-rings.
5279
Leak test result above al-
lowed limit
u
Resolve any leaks.
u
Clean and dry condenser
chamber.
u
Clean drain valve with water.
u
Clean O-rings.
5481
Info occurs while stoppering
step after the stoppering
vacuum is established.
u
Complete the manual
stoppering.
5570
Heatable shelf is discon-
nected.
u
Connect a functioning shelf at
the same connector position.
u
Defective component (e.g.
heating coil, cable, plug)
5571
Sample temperature sensor
is disconnected.
u
Connect a functioning PT-1000
at the same connector
position.
u
Defective component (e.g.
heating coil, cable, plug)
5704
System must be restarted af-
ter changing the network
settings
u
Restart the instrument.