10 | Cleaning and servicing
Büchi Labortechnik AG
88/106
Operation Manual K-365 Dist Line
u
Remove the decalcification solution from the instrument. See Chapter 12.4
"Removing water from the steam generator", page 98
u
Do steps (5) thru (10) again.
u
Install the H₂O supply for steam generation.
u
Set the On/Off master switch to On.
u
Press the
[READY]
button on the function bar.
u
Wait until the pumps stop working.
u
Set the On/Off master switch to Off.
u
u
Do steps (11) thru (16) five times.
10.16 Replacing the splash protector
NOTE
Installing is done in reverse sequence.
u
Remove the protection shield. See
Chapter 10.18 "Attaching and removing the
protection shield", page 89
u
Loosen the cap nut at the splash protector.
u
Open the splash protector nut.
u
Remove the splash protector.
u
Remove the steam connection.
u
Remove the NaOH/acid connection (according
to the instrument configuration).