10 | Cleaning and servicing
Büchi Labortechnik AG
86/106
Operation Manual K-365 Dist Line
10.14.
2
Calibrating the acid pump
Specifications on the display:
Precondition:
R
A sample tube is installed. See Chapter 8.9
"Installing a receiving vessel", page 73
u
Navigate to the specified submenu.
Pump Calibration
u
Select the specified action.
Acid
u
Tap the specified function on the function bar.
START
u
Wait until the display shows a numeric dialog.
...
u
Remove the sample tube.
u
Measure the volume in the sample tube.
u
Enter the value in the numeric dialog.
u
Tap the specified function on the function bar.
SAVE
10.14.
3
Calibrating the NaOH pump
Specifications on the display:
Precondition:
R
A sample tube is installed. See Chapter 8.9
"Installing a receiving vessel", page 73
u
Navigate to the specified submenu.
Pump Calibration
NaOH
u
Select the specified action.
...
Nominal Volume
u
Tap the specified function on the function bar.
EDIT
u
Enter the calibration volume you wish to use.
...
u
Tap the specified function on the function bar.
START
u
Wait until the display shows a numeric dialog.
...
u
Remove the sample tube.
u
Measure the volume in the sample tube.
u
Enter the value in the numeric dialog.
u
Tap the specified function on the function bar.
SAVE