Büchi Labortechnik AG
Cleaning and servicing | 10
Operation Manual K-365 Dist Line
83/106
Cleaning
If cleaning with a dishwashing machine:
u
Make sure, that there is no contact with other glassware.
Storage (e.g. drawer)
u
Prevent rolling and small hits.
ð
This can cause hair cracks which may lead to breakages.
10.3
Cleaning the glass components
The following values for the cleaning can be adjusted (the list differs according to the
instrument configuration):
[H₂O Volume]
[Steam Power]
[Distillation Time]
[Sample Tube Aspiration]
[Receiver Aspiration]
Specifications on the display:
Precondition:
R
A sample tube is installed. See Chapter 8.8
"Installing the sample tube", page 72
R
A sample receiving vessel is installed. See
Chapter 8.9 "Installing a receiving vessel",
page 73
u
Navigate to the specified submenu.
Cleaning
u
Adjust the values according to your needs.
...
u
Confirm the entries with the specified function
in the function bar.
SAVE
u
Tap the specified function on the function bar.
START
10.4
Cleaning and servicing the hoses and hose connections
u
Check the hoses and hose connections for defects (cracks, brittle areas).
u
If defective, replace the hoses.
10.5
Cleaning and servicing the sample tube seal
u
Check the seal for the following damage that may cause leakage:
Deterioration (surface roughness)
Deformation (rubber shape)
Residues (dirt, white crystals)
u
Rinse the seal with water.