3 Technical data
15
K-355
Operation Manual, Version G
3 .2
Technical data
Table 3-4: Technical data K-355
Distillation Unit K-355
Power consumption
max. 2.2 kW
Connection voltage
230 V ± 10% / 200 V ± 10%
Frequency
50/60 Hz
Current consumption (230 V)
8.5 A
Mains connection
3-pole (P, N, E) via power cord
Recovery rate
≥
99.5%
Reproducibility (RSD)
≤
1%
Detection limit
≥
0.1 mg Nitrogen
Environmental conditions
Temperature
Altitude
Humidity
for indoor use only
5 - 35 °C
up to 2000 m
maximum relative humidity 80% for temperatures up to
31 °C decreasing linearly to 67% relative humidity at
35 °C
Installation category
II
Pollution degree
2
Dimensions (W x H x D)
400 x 660 x 360 mm
Weight
17 kg
3 .3
Materials used
Table 3-5: Materials used for the K-355
Component
Material designation
Material code
Housing
Polyurethane
PUR / UL V0
Glass parts
Borosilicate glass 3.3
DIN/ISO 3585
Steam generator isolation
Ceramic fiber
Multitherm 550
Steam generator housing
Stainless steel
1.4301
Protective door
Polymethyl methacrylate
PMMA
Condenser cover
Polymethyl methacrylate
PMMA
Connecting stopper
Hypalon
CSM
NaOH tube
Ethylene Propylene Dimonomer
EPDM