6 | Operation
BÜCHI Labortechnik AG
50/102
Operation Manual Interface I-300 Pro
Fig. 26: Home screen after completion of distillation
u
After completion of distillation, press the AERATE function button.
ð
The system is vented until it reaches ambient pressure.
Manually stopping the cooling process
The cooling function continues after completion of a distillation process. If the system
has been pre-programmed accordingly, the cooling process stops after 5 minutes.
While cooling is active, the status bar shows the symbol
. The cooling process can
be stopped manually at any time regardless of how the system has been pre-pro-
grammed.
Precondition:
R
After completion of distillation, the system has been fully vented by pressing the
AERATE function button.
u
To manually stop the cooling process, press the function button
.
ð
The cooling process is stopped and the cooling symbol disappears from the status
bar.
Stopping everything
There is the possibility to immediately stopping all apparatus connected to the system
while distillation is in progress.
u
To stop all apparatus immediately, press the red STOP button (emergency stop).
6.2.3
Executing Manual mode
In
[Manual]
operating mode (manual distillation), the distillation process can be con-
trolled by manually setting the individual process parameters.
CAUTION
Risk of personal injury and property damage from unexpected equip-
ment behavior
u
Always carefully check the pre-programmed settings before every distillation
process.