BÜCHI Labortechnik AG
Installation | 5
Operation Manual Interface I-300 Pro
39/102
5.3
Connecting AutoDest sensor to vapor temperature sensor
(optional accessory)
The control unit offers programs for performing automatic distillation processes. The
programs require the connection of an AutoDest sensor. The AutoDest sensor is con-
nected to the inlet and outlet of the cooling condenser and continuously measures the
following temperatures:
Temperature of the incoming coolant
Temperature of the outgoing coolant
Temperature of the vapor exiting the evaporating flask
1
3
6
4
2
5
Fig. 16: AutoDest sensor, vapor temperature sensor and cooling condenser with evaporating
flask and receiving flask of a Rotavapor R-300
1
Vapor temperature sensor
4
Cooling condenser
2
AutoDest sensor
5
Coolant inlet on condenser
3
Communication connection between
AutoDest sensor and VacuBox
6
Coolant outlet on condenser
NOTE
There are two arrows stamped on the AutoDest sensor indicating the direction of flow
for the coolant. The coolant tubing should be connected accordingly.