Operation Manual UniversalExtractor E-800
17/88
3.1.4
Hot Extraction (with extraction glass chamber universal)
Step 1 extraction
The sample is located in the extraction chamber.
The beaker contains the solvent.
The solvent is heated, vapor rises up to the condenser, condenses
and drops into the extraction chamber with the sample.
The magnetic valve is closed, the solvent is collected up to the level
sensor.
The analyte is extracted.
The solvent is heated in the extraction chamber, vapor rises up to
the condenser, condenses and drops back into the extraction
chamber.
The magnetic valve opens regularly to release a small portion of
extract into the beaker.
Step 2 rinsing
The solvent is heated, vapor rises up to the condenser, condenses
and drops into the extraction chamber with the sample.
The magnetic valve is open, the solvent flows back into beaker, the
solvent is not collected.
Step 3 drying
The solvent is heated, vapor rises up to the condenser, condenses
and flows into tank.
The analyte remains in the beaker.