Formula for determining the daily factor
Determining the daily factor
Standard samples are analyzed to determine the daily factor. The known element concentration of the
standard samples is correlated with the actual element concentration computed.
Daily factor
The daily factor is computed according to this formula:
whereas:
f = daily factor
c
theo
= theoretical element concentration
c
act
= element concentration computed actually