3 Technical data
12
Melting Point M-560 Operation Manual, Version G
Table 3-4: Technical data of the Melting Point M-560
Melting Point M-560
Accuracy of boiling point at 1.0 °C/min up to 400 °C
±
0.5 °C
Repeatability of boiling point at 1.0 °C/min
±
0.3 °C
Temperature gradients, °C/min
0.1, 0.2, 0.5, 1, 1.5, 2, 2.5, 3, 5, 10, 20
Heat-up time (50 °C–350 °C) at 25 °C
~4 min
Cool-down time (350 °C–50 °C) at 25 °C
~13 min
Electrical supply
100–240 V (±10 %), 50–60 Hz
Power consumption
150 W
Contact termination
L, N, PE
Approval
CE, CSA, UL
Dimensions (W×H×D), mm
190×200×370
Weight, kg
4.5
Environmental conditions
For indoor use only
Temperature
5–40 °C
Altitude
up to 2000 m a.s.l.
Humidity
Maximum relative humidity 80 % for temperatures up to
31 °C, decreasing linearly to 50 % relative humidity at 40 °C
Overvoltage category
II
Degree of protection
IP20
Pollution degree
2
Storable methods for melting point
50
Storable methods for boiling point
50
Compliant with Pharmacopeia methods
PH. EUR., USP and JP
NOTE
Temperature measuring accuracy refers to pharmacopoeia melting point.
3.3
Materials used
Table 3-5: Materials used
Component
Material designation
Print holder
PA
Heating block
Aluminium
Lenses
Glass
Axial fan
Aluminium
Housing
PU, Stainless Steel, Glass
Cover
POM, Ceramic, Aluminium, Stainless Steel
OM_093251_M-560_en_G.indd 12
12.07.2018 09:21:14