T419S Indicator Service Manual
35
6.2
Scale menu
6.2.1
Capacty
This stands for capacity. Access this item and you can key in a new capacity or accept
the displayed capacity.
6.2.2
Dvision
This stands for division size. Choices are
0.000001
,
0.000002
,
0.000005
,
0.00001
,
0.00002
,
0.00005
,
0.0001
,
0.0002
,
0.0005
,
0.001
,
0.002
,
0.005
,
0.01
,
0.02
,
0.05
,
0.1
,
0.2
,
0.5
,
1
,
2
,
5
,
10
,
20
,
50
,
100
,
200
and
500
. The default value is
1
.
6.2.3 Units
Use this to select the units of measure or create custom units.
Unit
From this item you can choose which units of measure are available when
the user presses the
UNITS
key during normal weighing. You can have up
to four units for viewing. They are listed as:
Unit 1
,
Unit 2
,
Unit 3
and
Unit
4
. You can assign any of the following units of measure to any of the units:
Lb
,
1000 g
,
oz
,
gr
,
lb-oz
,
Cust
1
,
Cust
2
,
Cust
3
,
Cust 4
or
Off
.
Custom
This stands for custom unit. To use a custom unit you must first create
them by entering a ratio and a string to define it.
Cust 1-4
You can set up one to four custom units. The process to set
up a custom unit is the same for all four.
Ratio
The ratio is the number you divide into the
calibration unit of measure to create the custom
unit. Example: Ratio would equal 2000 if you
wanted to convert pounds to tons.
String
Use this to enter a string label for the custom unit.
6.2.4 Stable
Use this parameter to set the stability window for the scale. Set a division window and
a time window which will be used to determine when the stability icon will be displayed.
Div
Set the division window size to define stability.
Time
Set the time window in seconds to define stability. Set both
Div
and
Time
to
0
to disable stability.
Division sizes for other units of measure are automatically calculated by the indicator.
The
Stable
time value is used during normal operation as the length of time the
indicator will continue to check for a ‘motion stable’ condition after the
ZERO
,
TARE
or
button is pressed. If the intended operation cannot be completed before the
timeout,
Cant
is displayed and the key request is ignored.
Summary of Contents for T419S
Page 2: ...BTEK_T419S_s_en_501788 book 2019 B TEK All rights reserved ...
Page 118: ...118 T419S Indicator Service Manual ...
Page 130: ...130 T419S Indicator Service Manual ...
Page 134: ...134 T419S Indicator Service Manual Last 28 Temp 28 ...
Page 135: ......
Page 136: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088 ...