Values Tab
The
Values
tab displays critical values pertaining to the current test, such as information for the current tracker under
test.
MPE coefficients
and
Uncertainty Budget
terms are also displayed, and can be modified here (for the current test
only). This functionality allows you to make test-specific edits rather than use default values if you desire. As an
example, you might alter the uncertainty budget value for the
retroreflector
when using targets of various grades.
Alternatively, you may wish to verify your laser tracker to a different standard than that reported by the manufacturer.
Changes to MPE coefficients or uncertainty budget items on the
Values
tab will be reflected in the error graph on the
Results