51
Z31914_00_01
Operation
10.1.4.4
Nitrogen/Air Flow Information
The nitrogen or air flow are used on cooled racks to keep samples from freezing. The
Nitrogen/Air Flow Information panel displays a target flow value and the current flow.
The value fields are color coded, when the flow value is within the target flow value, the
Current Flow field border is green.
Figure 10.8 Nitrogen/Air Flow Information
When the difference in the target and current values exceed preset values, an error
message appears and the current flow field border will turn to red. These preset values
must be entered by a trained service representative.
10.1.4.5
Cooling Temperature Information
The Cooling Temperature Information panel display the target and current temperatures
of the selected rotor container, as well as the rotor container ID.
Figure 10.9: Cooling Device Information
The value fields are color coded, when the temperature value is within the target temper-
ature value the current temperature field border is green.
When the difference in the target and current values exceed preset values, an error
message appears and the Current Temperature field border will turn to red. These pre-
set values must be entered by a trained service representative.
Summary of Contents for SamplePro hr-MAS
Page 6: ...6 Z31914_00_01 Table of Contents ...
Page 38: ...38 Z31914_00_01 Installation and Initial Commissioning ...
Page 44: ...44 Z31914_00_01 Assembling the Tray with Rotors ...
Page 57: ...57 Z31914_00_01 Operation Figure 10 14Activity History ...
Page 58: ...58 Z31914_00_01 Operation ...
Page 62: ...62 Z31914_00_01 Preventative Maintenance ...
Page 65: ...65 Z31914_00_01 13 EC Declaration of Conformity ...
Page 66: ...66 Z31914_00_01 EC Declaration of Conformity ...
Page 68: ...68 Z31914_00_01 Contact ...
Page 70: ...70 Z31914_00_01 Warning Signs ...
Page 72: ...72 Z31914_00_01 Figures ...
Page 74: ...74 Z31914_00_01 Tables ...
Page 76: ...76 Z31914_00_01 Glossary ...
Page 80: ...80 Z31914_00_01 Index ...
Page 81: ......