51
ZTKS0156 / Z31820 / Rev.: 03
Assembling
4.3.2.2
Mounting the Vacuum Valve at Evacuated Cryostat
i
The sealing plug can not be moved until a correct vacuum was applied at the KF 40
flange at the Vacuum Valve.
Figure 4.28: Mounting the Vacuum Valve at evacuated cryostat
1.
Remove the protection cap of the vacuum
flange at the bottom plate (2) of the cryostat.
2.
Install the valve stem (10) onto the sealing
plug (3) and tighten it slightly.
3.
Turn the Vacuum Valve in the desired posi-
tion (KF 40 flange (9) looking outwards).
4.
Place the two half rings (7) as shown into
the slots of the valve body (1).
5.
Fix the half rings (7) with four M 6 x 12 (8)
screws.