101
ZTKS0156 / Z31820 / Rev.: 03
Troubleshooting
6.2.7
During De-energizing and Warming up
Indicator
Possible reason
Solution
By
The magnet system
quenches during de-
energizing.
The helium level was
too low for de-ener-
gizing.
Refill helium at least to the
minimum allowed level (see
Appendix Technical Data).
Bruker Service
The power supply is
defective.
Replace power supply.
Bruker Service
The main current is
oscillating.
Replace power supply.
Bruker Service
The shim current
can not be set
correctly.
The control cable is
not connected cor-
rectly to the current
lead and/or the
power supply.
Connect the control cable to
the current lead and to the
power supply correctly.
Bruker Service
The switch "Main
Coil/OFF/Shim Coil"
is not on the "Shim
Coil" position.
Switch "Main Coil/OFF/Shim
Coil" on the "Shim Coil" posi-
tion.
Bruker Service
High helium flow
after breaking
vacuum.
Remaining cryo-
genic agents in the
inner vessels.
Remove liquid helium.
Bruker Service
Vacuum still remains
after 12 hours.
Vacuum valve is
closed.
Open vacuum valve. Block it if
necessary.
Bruker Service
RT vessel is wet and
cold.
Cryostat is still cold.
Wait until RT vessel is dry and
warm.
Check PT 100 temperature
sensors.
Bruker Service
RT bore wet and
cold before
disassembling.
Cryostat is still cold.
Wait one more day.
Do not open a cryostat before
the room temperature bore
tube is warm and dry!
Bruker Service
Summary of Contents for RS
Page 1: ...NMR Spectroscopy Service Manual RS ...
Page 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Page 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Page 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Page 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Page 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Page 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Page 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Page 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Page 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Page 156: ...156 ZTKS0156 Z31820 Rev 03 ...