Cabinet Safety
18 / 48
H171764_6_001
Lifting the AQS Chassis
WARNING: At least two people are needed to insert and remove the AQS chassis from the
electronics cabinet. A fully equipped AQS system can weight in excess of 50kg.
NOTE: Remove some or all of the AQS units from the chassis prior to handling to reduce
weight.
Cleaning
WARNING: Always switch power off and disconnect the power cable before cleaning. Never
power on until all surfaces are completely dry.
Clean the outside of the AQS chassis and units with a soft, lint-free cloth dampened in water.
Do not use any detergent or other cleaning solvents.
3.5
BSMS/2
The enhanced BSMS/2 system contains a set of highly integrated boards (ELCB and
SCB20), which provide optimal performance, higher resolution and increased stability. The
BSMS/2 mainframe is designed as a subunit in the electronics cabinet of the NMR
Spectrometer. For the environmental conditions please refer to the site planning guide of the
spectrometer system.
The following general safety precautions must be observed during all phases of operation
and service of the BSMS/2 system. Failure to comply with these precautions or with specific
warnings elsewhere in this manual violates safety standards of design, manufacture and
intended use of the BSMS/2 system.
BRUKER assumes no liability for the customer‘s failure to comply with these requirements
and is therefore not responsible or liable for any injury or damage that occurs as a
consequence of non-approved manipulations on the BSMS/2 system.
3.5.1
Emergency Shutoff
In case of an emergency, use the mains switch described in the console
to switch off the complete console. The BSMS in AVANCE NEO systems can only be
switched off through the PDU. It is not advised to selectively switch off the BSMS. Instead,
shutdown the console completely. Only if otherwise instructed by Bruker personnel, you may
be guided through a process to selectively shutdown the BSMS.
3.5.2
Personnel Safety
Ground Connection
WARNING: To minimize shock hazard the BSMS/2 chassis must be connected to an
electrical ground.
The electronics cabinet is equipped with a three-conductor AC power cable. Do only use
power cables approved by BRUKER or compliant with IEC safety standards.
Summary of Contents for NMR AVANCE NEO Systems
Page 40: ...Contact 40 48 H171764_6_001 ...
Page 42: ...List of Figures 42 48 H171764_6_001 ...
Page 44: ...List of Tables 44 48 H171764_6_001 ...
Page 46: ...Index 46 48 H171764_6_001 ...
Page 47: ...H171764_6_001 47 48 ...
Page 48: ...Bruker Corporation info bruker com www bruker com Order No H171764 ...