Technical Data
16
H148851_4_004
4.3
Gas Connection
One fitting is provided for nitrogen purge gas. The nitrogen source must be regulated to a
pressure of no more than 35 kPa (5 psi, 0.35 bar). A typical gas flow of 0.5 liter/minute shall
apply.
4.4
Thermal Characteristics
Temperature calibration and validation is performed at the factory.
Figure 4.1: Thermal Characteristics of a 3-Zone 10 mm SFC Version
Zone
Default Temperature
Temperature Range
3
0° C
0° C to +66° C
2
65° C
A5° C to +66° C
1
100° C
A5° C to +100° C