Introduction
30
E1400912_1_001
WARNING
Danger of injury due to falling lid
The hinged lids may close unexpectedly during probe exchange causing injury to hands or
fingers.
u
Ensure that the gas spring on the outer lid is operating correctly.
u
The blue inner hinged lid does not have a gas spring, use extreme care when
exchanging the probe assembly.
4. The probe assembly is fixed with two quick-release fasteners. To release the fasteners,
lift up the D-rings until they snap into a vertical position, then use them to rotate the
fastener posts until released (refer to the figure below).
5. While grasping the D-rings, carefully pull upward to loosen the preamplifier connection
under the mounting plate, than raise the assembly (mounting bar with probe and
preamplifier attached) out of the magnet gap.
Figure 5.6: Ring of the Quick-Release Fastener (left) and the Raising of the Probe Assembly (right)
1.
Rings
2.
Rotating the rings.
To open the fasteners, turn the rings (1) upward and rotate them (2)
Replace the assembly by following the instructions in reversed order.
Intelligent
probes were developed for the minispec mq-Series. When you exchange a probe,
the software will read the most important parameters (pulse length, resonance frequency)
from the probe itself. The software will prompt the user to confirm the settings. After
temperature stabilization it is recommended to run the Daily Check procedure. In certain
cases a fine tuning by running Update Settings may be beneficial.
Summary of Contents for minispec mq20
Page 1: ...minispec mq20 Toothpaste Analyzer User Manual Version 001 Innovation with Integrity AIC ...
Page 5: ...Contents E1400912_1_001 5 Index 83 ...
Page 6: ...Contents 6 E1400912_1_001 ...
Page 14: ...Safety 14 E1400912_1_001 ...
Page 18: ...Transport Packaging and Storage 18 E1400912_1_001 ...
Page 22: ...Technical Data 22 E1400912_1_001 ...
Page 72: ...minispec exe Software for Diagnostic and Service Purposes 72 E1400912_1_001 ...
Page 76: ...Maintenance 76 E1400912_1_001 ...
Page 78: ...Dismantling and Disposal 78 E1400912_1_001 ...
Page 80: ...Contact 80 E1400912_1_001 ...
Page 82: ...List of Figures 82 E1400912_1_001 ...
Page 85: ...E1400912_1_001 85 ...
Page 86: ...Bruker Corporation info bruker com www bruker com Order No E1400912 ...