48
IFS 125M User Manual
Bruker Optik GmbH
Operation 5
5.6
Checking the signal
Especially after you have replaced a spectrometer component (source, laser, beamsplit-
ter, detector), it is advisable to check whether a signal is detected, and to check the sig-
nal intensity (signal amplitude).
If a liquid nitrogen-cooled detector is used, the detector must be cooled (chapter 5.9)
before a signal can be detected.
5.6.1
Procedure
1.
Make sure that the spectrometer sample compartment is empty, or contains a gas
cell the windows of which are transparent in the spectral range desired.
2.
On the OPUS
Measure
menu, select the
Advance Measurement
command.
3.
Select appropriate measurement parameters.
4.
Click the
Check Signal
tab.
5.
Activate the
Interferogram
option button.
Figure 5.1:
OPUS - Check Signal tab
B
A
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...