172
IFS 125M User Manual
Bruker Optik GmbH
Data Communication 9
9.1
Configuration types
9.1.1
General rules
• All IP addresses must be unique within a given network. Never assign the same
IP address, e.g. to the spectrometer and the computer.
• The spectrometer can only be configured using static IP addresses (no DHCP
capability).
• The IP addresses of the spectrometer and the PC must be within the same sub-
net, as long as no gateway is used.
9.1.2
Connecting spectrometer to a stand-alone PC (standard)
1.
Connect one end of the supplied cross-over data cable to the ETH port on the
spectrometer rear side.
Figure 9.1:
RJ-45 Ethernet cable wiring
Straight-through cable
Cross-over cable
Figure 9.2:
Connecting spectrometer to stand-alone PC