132
IFS 125M User Manual
Bruker Optik GmbH
Troubleshooting 7
i
Data acquisition can be tested independent of any detector signal by sending the direct
command:
TRW=1
. In that case a triangular wave signal is recorded instead of the digi-
tized detector signal. To return to normal operation, send the direct command:
TRW=0.
Detector oversaturation or
ADC overflow
• Reduce the light source intensity by using a
smaller aperture.
• Reduce the gain settings. On the OPUS
Measure
menu, select the
Advanced Measure-
ment
command. Click the
Optics
tab.
Scanner not scanning
Check the FWD and TKD LEDs on the mobile
electronics unit. If they do not light repeatedly,
proceed as follows:
• On the OPUS
Measure
menu, select the
Optic
Setup and Service
command.
• Click the
Control Panel
tab on the dialog that
opens.
• Scroll to the
Scanner control
item.
• Click once on
Reinit Scanner
. After a while the
FWD and TKD LEDs on the mobile electronics
unit flash, indicating that the scanner is in short
scan mode.
Interferometer out of align-
ment
Contact Bruker service.
Possible causes
Solutions
Table 7.4: No interferogram observable
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...