91
Bruker Optik GmbH
HYPERION User Manual
Operation 4
4.13
Measuring a sample with the ATR objective (single point
measurement)
Operating step
Corresponding operating element(s)
1. Activate the viewing mode.
either the VIS button on the operating
panel (B in fig. 3.4) or in the OPUS video
wizard
☞
See OPUS/VIDEO Manual.
2. Activate the reflection mode.
either the reflection mode button on the
operating panel (E in fig. 3.4) or in the
OPUS video wizard
☞
See OPUS/VIDEO Manual.
3. Put the sample on the stage.
4. Swing the ATR objective in the
beam path.
Bring the ATR crystal in the viewing
mode position (i.e. upper position).
Set a contact pressure level which is
adequate to the degree of hardness of
the sample you intend to analyze.
ATR objective
☞
See section 4.12.4 and section
4.12.5
5. Set the microscope either for sam-
ple viewing through the binocular or for
video image display.
Note: It is recommended to use the
video image displayed on the monitor.
Light path selector lever (A in fig 3.14)
6. Move the stage in x- and/or y-direc-
tion to find a sample area which is of
interest to you.
Manual stage:
Rotary knobs (A and B in
fig. 3.6)
Motorized stage:
Joystick and corre-
sponding stage control functions in
OPUS
☞
See OPUS/VIDEO Manual.
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...