41
Bruker Optik GmbH
HYPERION User Manual
Overview 3
3.7.4
Iris aperture for setting the Koehler illumination
In reflection mode
This aperture (E in fig. 3.10) is used for setting the Koehler illumination in reflection
mode. The illumination intensity can be set in a stepless manner.
In transmission mode
This aperture (B in fig. 3.10) is used for setting the Koehler illumination in transmission
mode. It is located underneath the microscope stage. This aperture is opened and
closed by rotating the knurled ring.
B
Lever in front position:
When the lever is in the front position, the aperture can be
rotated by max. 360° using the thumbwheels (A in fig. 3.11). In this case, the previ-
ously set aperture opening size is locked and the knife-edge aperture is rotated. The
rotation center is identical with the center of the viewing field.
C
Lever in back position:
When the lever is in the back position, the distance between
two opposite knife edges (i.e. the size of the rectangular aperture opening) can be
increased or reduced using the two thumbwheels (A in fig. 3.11). In doing so, the size
of the rectangular aperture opening is set.
Fig. 3.11
Operating elements of the knife-edge aperture
Figure 3.12: Operating element of the aperture for setting Koehler illumination in reflection mode
Lever
Lever position
Aperture setting
Lever is pushed in completely. Aperture
is
open.
Lever is pulled out completely.
Aperture is closed.
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...