30
HYPERION User Manual
Bruker Optik GmbH
Overview 3
3.4
Brightness control and brightness indicator
G
Detector temperature warning indicator:
lights up red when the MCT detector tem-
perature exceeds its operating temperature. In case the microscope is equipped with
two MCT detectors, a red point lights up left or right in the display depending on
which detector has exceeded its operating temperature.
Note: In this case, the detector needs to be cooled with liquid nitrogen. For informa-
tion about how to cool the detector, see section 4.4.
H
Start button:
to start a spectroscopic measurement directly from the microscope
using the current measurement parameter settings in OPUS.
Note: Before pressing this button, make sure that either the IR button or VIS/IR but-
ton is activated, the OPUS software program is open and appropriate measurement
parameter settings are selected. (See appendix B in this manual and the OPUS Ref-
erence Manual.)
Fig. 3.4
Control element and explanation
Figure 3.5:
Adjusting the brightness intensity of the microscope illumination
A
B
C
Fig. 3.5
Control element and explanation
A
Brightness indicator:
indicates the current brightness intensity of the
visible illumination in relation to the maximum possible brightness inten-
sity by means of a bar graph display.
B
Brightness control:
Thumb wheel for adjusting the brightness intensity
of the microscope illumination in a continuous manner.
C
Icon legend
to illustrate how to increase / decrease the brightness inten-
sity
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...