Lock Error Messages
100
Spectrospin AG
Daedalus Lock Manual
LOCK Wrongaddress
Error 32
The lock application software is a INTEL Hex file. If any
address in this file is bigger than the uController(80C535)
serves with its addressbus(0xFFFF) downloading will stop
immediately.
Your download file is damaged, contact your local Bruker
service agent for a correct LOCKxx.HEX file.
To restart the download, reset the BSMS and clear the LOCK
Errormessage #11.(The lock uses only the booteprom, no
valid application software in its program memory).
LOCK Wrong AppSW
Error 12
The lock boots from an EPROM. The bootroutine stored in
this EPROM gets an lock specific identifier. While the lock
powers up, the boot identifier is compared to the application
software identifier. The ”download all boards” function in the
BSMS service tool guarantees software consistency. To
restart downloading, take out jumper 6, reset the BSMS, clear
the lock errormessage #11 and start “download all boards“.
After the download has been done, the lock remains in
errorstate #11. Put in jumper 6 and reset the BSMS once
again.
LOCK WrongRecord Type
Error 33
The lock application software is a INTEL Hex file. If any line
record type specifier violates the format specifications
downloading will stop immediately.
Your download file is damaged, contact your local Bruker
service agent for a correct LOCKxx.HEX file.
To restart the download, reset the BSMS and clear the LOCK
Errormessage #11.(The lock uses only the booteprom, no
valid application software in its program memory).
LOCK WrongCheckSum
Error 34
The lock application software is a INTEL Hex file. Every line
in this file gets a checksum. The lock calculates for each line
its own checksum and compares them. If a checksum differs
from the calculated checksum downloading will stop
immediately.
Your download file is damaged, contact your local Bruker
service agent for a correct LOCKxx.HEX file.
To restart the download, reset the BSMS and clear the LOCK
Errormessage #11.(The lock uses only the booteprom, no
valid application software in its program memory).
LOCK WrongTransmissionCheck
Error 38
The lock application software is a INTEL Hex file. A stopline
indicates the end of file. If this end of file line violates the file
specification downloading will stop immediately.
Your download file is damaged, call your local Bruker service
agent for a correct LOCKxx.HEX file.
To restart the download, reset the BSMS and clear the LOCK
Errormessage #11.(The lock uses only the booteprom, no
valid application software in its program memory).
Table 12. Lock Error Messages in alphabetical order
ERROR MESSAGE
ERROR DESCRIPTION
SIGNALS
Note: Signals of
the serial bus
(power, gain
phase, shift etc.)
are only active
during a process
change
Summary of Contents for DAEDALUS
Page 1: ...BRUKER BRUKER SPECTROSPIN AG SERVICE MANUAL DAEDALUS LOCK Version BSMS 004 ...
Page 6: ...Contents 6 Spectrospin AG Daedalus Lock Manual ...
Page 14: ...8 Spectrospin AG Daedalus Lock Manual ...
Page 16: ...10 Spectrospin AG Daedalus Lock Manual ...
Page 26: ...General Description 20 Spectrospin AG Daedalus Lock Manual ...
Page 36: ...Operation 30 Spectrospin AG Daedalus Lock Manual ...
Page 56: ...Lock Transmitter 50 Spectrospin AG Daedalus Lock Manual ...
Page 64: ...Lock Receiver 58 Spectrospin AG Daedalus Lock Manual ...
Page 68: ...19F RX Option 62 Spectrospin AG Daedalus Lock Manual ...
Page 85: ...Installation Daedalus Lock Manual Spectrospin AG 79 Installation 11 2 ...
Page 86: ...Z0 Compensation Option 80 Spectrospin AG Daedalus Lock Manual ...
Page 90: ...Technical Data 84 Spectrospin AG Daedalus Lock Manual ...
Page 92: ...Trouble Shooting 86 Spectrospin AG Daedalus Lock Manual ...
Page 98: ...BSMS service tool 92 Spectrospin AG Daedalus Lock Manual ...
Page 107: ...Error Descriptions Daedalus Lock Manual Spectrospin AG 101 ...
Page 108: ...Lock Error Messages 102 Spectrospin AG Daedalus Lock Manual ...
Page 114: ...Appendix 108 Spectrospin AG Daedalus Lock Manual ...
Page 122: ...116 Spectrospin AG Daedalus Lock Manual ...
Page 123: ...Daedalus Lock Manual Spectrospin AG 117 Lock Transmitter Index B ...
Page 124: ...118 Spectrospin AG Daedalus Lock Manual ...
Page 125: ...Daedalus Lock Manual Spectrospin AG 119 Lock Transmitter Index C ...
Page 126: ...120 Spectrospin AG Daedalus Lock Manual ...
Page 127: ...Daedalus Lock Manual Spectrospin AG 121 19F Transmitter Option Index A ...
Page 128: ...122 Spectrospin AG Daedalus Lock Manual ...
Page 129: ...Daedalus Lock Manual Spectrospin AG 123 Lock Receiver Index C ...
Page 130: ...124 Spectrospin AG Daedalus Lock Manual ...
Page 131: ...Daedalus Lock Manual Spectrospin AG 125 Lock Receiver Index D E ...
Page 132: ...126 Spectrospin AG Daedalus Lock Manual ...
Page 133: ...Daedalus Lock Manual Spectrospin AG 127 19F Receiver Option Index A ...
Page 134: ...128 Spectrospin AG Daedalus Lock Manual ...
Page 135: ...Daedalus Lock Manual Spectrospin AG 129 Lock Controller Index A ...
Page 136: ...130 Spectrospin AG Daedalus Lock Manual ...
Page 137: ...Daedalus Lock Manual Spectrospin AG 131 Lock Controller Index B C ...
Page 138: ...132 Spectrospin AG Daedalus Lock Manual ...
Page 139: ...Daedalus Lock Manual Spectrospin AG 133 Lock RS232 Biggy Board ...
Page 140: ...134 Spectrospin AG Daedalus Lock Manual ...
Page 141: ...Daedalus Lock Manual Spectrospin AG 135 Z0 Compensation Option ...
Page 142: ...136 Spectrospin AG Daedalus Lock Manual ...
Page 143: ...Daedalus Lock Manual Spectrospin AG 137 Cables ...
Page 144: ...138 Spectrospin AG Daedalus Lock Manual ...
Page 145: ...Daedalus Lock Manual Spectrospin AG 139 Test Equipment Index A ...