Alarm System
The alarm system is an important part of the BMPC/2 and is needed for the supervision
of the magnet system. The alarm system uses an analog telecommunication network,
based on an analog fax modem (with its own modem controller), and normally is avail-
able for use worldwide.
6.1
Alarm System Function
When the BMPC/2 is working in the Monitor & Alarm mode, all system parameters are
continuously monitored and compared with their limit values. When a parameter
exceeds or falls below a given alarm limit, an alarm cycle will be started.
In the following example the alarm cycle was triggered because of a low helium flow.
The status field shows that an alarm call was activated and the resulting events logged
in the Event Log section.
Figure 6.1 Present Values/Events in Alarm Mode
1. Helium Flow Alarm is enabled
2. Alarm Fax is sent to Bruker BioSpin
3. Confirmed by „Mustermann, Klaus“
4. System called Mustermann, Klaus
5. Helium Flow is below alarm limit
6. System must be in Monitor & Alarm Mode
7. System status is now Alarm Call
7
6
5
5
4
3
2
1
Summary of Contents for BMPC/2
Page 8: ...8 Z31823_00_05 Table of Contents...
Page 14: ...14 Z31823_00_05 Safety...
Page 20: ...20 Z31823_00_05 Principles of Operation...
Page 42: ...42 Z31823_00_05 Alarm System...
Page 62: ...62 Z31823_00_05 Hardware Overview...
Page 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Page 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Page 122: ...122 Z31823_00_05 Technical Data...
Page 124: ...124 Z31823_00_05 Contact...
Page 126: ...126 Z31823_00_05 Warning Signs...
Page 130: ...130 Z31823_00_05 Figures...
Page 132: ...132 Z31823_00_05 Tables...
Page 135: ......